Effect of evacuated blood collection tubes on thromboplastin calibration.
van den Besselaar AM, Meeuwisse-Braun J, Witteveen E, van Meegen E
Purpose
The purpose of this paper was to determine the effects of collection container on thromboplastin calibration.
Conclusion
Mean prothrombin time (PT) was dependent on both the collection tube type, and coagulometer. However the magnitude of the effects of collection container and coagulometer were dependent on the PT such that the ratio of patient PT to normal PT for each collection container and coagulometer was different.
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection tube and coagulometer on PT. Patients on anticoagulant therapy and normal individuals were examined.
Summary of Findings
Regardless of the instrument used, mean PT for normal individuals and patients on anticoagulant therapy was longest when specimens were collected in plastic Vacuette tubes and shortest when collected in the glass Venoject tubes. Average PT was shorter when measured using the KC-10 coagulometer versus the MLA Electra 1800. Collection container significantly influenced the ratio of PT of patients on oral anticoagulant therapy to that in normal individuals. The largest difference in this ratio was observed between plastic Vacuette and the glass Venoject samples when measured on the MLA Electra 1800. The smallest differences in this ratio were between the glass Vacutainer and Venoject tubes. The authors note that some differences in PT may be attributable to different concentrations of citrate in the tubes in addition to tube material.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Other drugs Long term anticoagulant therapyNoneBiospecimen Acquisition Type of collection container/solution Plastic VacuetteSilicone coated glass VacutainerSilicone coated glass VenojectHematology/ auto analyzer Specific Technology platform KC-10MLA Electra 1800