Skip to main content

A comparison of blood collection in glass versus plastic vacutainers on results of esoteric coagulation assays

Flanders MM, Crist R, Rodgers GM

Publication: Lab Med (2003) • Vol 34 • Page 732

Purpose

The purpose of this paper was to determine the effects of collection container material on hematological parameters of normal sodium citrate plasma specimens.

Conclusion

Plasma specimens collected in plastic vacutainers had increased thrombin time (TT), a slight increase in protein S antigen and activity, and a small decrease in Factors XII, VII and VIII compared to specimens collected in glass vacutainers. However, only the increased TT when plasma was collected in plastic instead of glass was large enough to be clinically significant.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection container material on hematological parameters of normal sodium citrate plasma specimens. Specimens were drawn sequentially into glass vacutainers first and then into plastic vacutainers.

Summary of Findings

The correlation between glass and plastic vacutainers was greater than 0.85 for all assays except TT. Plasma specimens collected in plastic vacutainers had increased TT, a slight increase in protein S activity, and a small decrease in factors XII, VII and VIII compared to specimens collected in glass vacutainers. However, only the increased TT when plasma was collected in plastic instead of glass was large enough to be clinically significant. Collecting specimens into plastic instead of glass vacutainers had no effect on the activity or antigen of Protein C, antithrombin, vonWillebrand factor, ristocetin cofactor, activated protein C resistance, fibrinogen activity, or coagulation factors II, V, IV, X or XI.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinELISA
    ProteinHematology/ auto analyzer
    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass vacutainer
    Plastic vacutainer