A new plastic evacuated tube with plasma separator.
Landt M, Wilhite TR, Smith CH
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine the effects of tube type, and centrifugation speed (1100 x g or 13600 x g) and duration (1, 2, 3, 5 or 10 min) on platelet removal, lactate dehydrogenase (LDH) levels and plasma recovery.
Summary of Findings
Centrifugation of blood at 13,600 x g reduced platelet counts in the plasma by more than 95% after only 3 minutes and resulted in maximum plasma recovery after just 1 minute. When blood was centrifuged in plastic tubes at 1100 x g, 10 minutes was necessary to effectively remove platelets and recover the maximum amount of plasma. The amount of plasma recovered at each time point was dependent on the hematocrit of the specimen. No effect of centrifugation or tube type was found on LDH.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Cell count/volumeCoulter counterProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level Hematocrit of 35%Hematocrit of 46%Hematocrit of 58%Biospecimen Acquisition Type of collection container/solution GlassTerumo Venoject IIBiospecimen Aliquots and Components Blood and blood products PlasmaWhole bloodBiospecimen Aliquots and Components Centrifugation Multiple durations comparedMultiple speeds compared
Purpose
Summary of Findings
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
Small moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution GlassTerumo Venoject IIBiospecimen Aliquots and Components Aliquot size/volume 1 mL2 mL4 mLStorage Storage duration 0 h24 hStorage Storage temperature 4 degrees CStorage Type of storage container Plastic with separatorPlastic without separator