Effects of needle size and storage temperature on measurements of serum potassium.
Verresen L, Lins RL, Neels H, De Broe ME
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of needle gauge and storage of blood specimens on serum potassium levels.
Summary of Findings
The average potassium concentration was lower when blood was drawn through a 19 gauge needle than when drawn through a 23 gauge needle (p<0.001). Serum potassium concentrations increased when clotted blood was stored for 16 h at 4 degrees C (p<0.001), however no change in potassium levels occurred when clotted blood was stored for 16 h at 18 degrees C compared to that which was centrifuged and analyzed immediately.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Needle gauge 1923Biospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 h16 hStorage Storage temperature 18 degrees C4 degrees C