Effects of needle size and storage temperature on measurements of serum potassium.

Verresen L, Lins RL, Neels H, De Broe ME

Publication: Clin Chem (1986) • Vol 32 • Page 698-9PMID: 3955823View on PubMed

Purpose

The purpose of this paper was to determine the effects of needle gauge and storage of blood specimens on potassium levels.

Conclusion

The average potassium concentration was lower when blood was drawn through a 19 gauge needle than when drawn through a 23 gauge needle. Potassium concentrations increased when blood was stored for 16 h at 4 degrees C, however no change in potassium levels occurred when blood was stored for 16 h at 18 degrees C.

Studies (1)

Purpose

The purpose of this study was to determine the effects of needle gauge and storage of blood specimens on serum potassium levels.

Summary of Findings

The average potassium concentration was lower when blood was drawn through a 19 gauge needle than when drawn through a 23 gauge needle (p<0.001). Serum potassium concentrations increased when clotted blood was stored for 16 h at 4 degrees C (p<0.001), however no change in potassium levels occurred when clotted blood was stored for 16 h at 18 degrees C compared to that which was centrifuged and analyzed immediately.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionNeedle gauge
    19
    23
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 h
    16 h
    StorageStorage temperature
    18 degrees C
    4 degrees C