Hemolysis in needleless connectors for phlebotomy.
Sharp MK, Mohammad SF
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of needle gauge, the use of needleless connectors, and different collection containers on hemolysis. All blood draws simulated venipuncture by drawing blood from a bottle through the needles or connectors into a vacutainer tube or syringe.
Summary of Findings
The use of needleless connectors resulted in more hemolysis than blood collection using needles of any gauge, but regardless of collection method, hemolysis was low (<43 mg/dL) and unlikely to interfere with clinical assays. Specimens collected using the PosiFlow needless connector had more hemolysis than those collected with the Clave needleless connector (p=0.0008). Of the needles, use of the 18 gauge needle resulted in the most hemolysis and the 22 gauge needle resulted in the least (p=0.008). Using a vacutainer instead of a syringe for blood collection slightly increased hemolysis (p=0.023).
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition NeedleNeedleless connectorSyringe drawVacuum tubeBiospecimen Acquisition Needle gauge 161822Biospecimen Acquisition Type of collection container/solution SyringeVacutainer tube