Hemolysis in needleless connectors for phlebotomy.

Sharp MK, Mohammad SF

Publication: ASAIO J (2003) • Vol 49 • Page 128-30PMID: 12558320View on PubMed

Purpose

The purpose of this paper was to determine the effect of drawing blood with needleless connectors on hemolysis.

Conclusion

As hemolysis was low (<43 mg/dL) in all specimens, regardless of needle gauge or needleless connector, the authors believe the use of needless connectors is unlikely to interfere with common clinical assays.

Studies (1)

Purpose

The purpose of this study was to determine the effects of needle gauge, the use of needleless connectors, and different collection containers on hemolysis. All blood draws simulated venipuncture by drawing blood from a bottle through the needles or connectors into a vacutainer tube or syringe.

Summary of Findings

The use of needleless connectors resulted in more hemolysis than blood collection using needles of any gauge, but regardless of collection method, hemolysis was low (<43 mg/dL) and unlikely to interfere with clinical assays. Specimens collected using the PosiFlow needless connector had more hemolysis than those collected with the Clave needleless connector (p=0.0008). Of the needles, use of the 18 gauge needle resulted in the most hemolysis and the 22 gauge needle resulted in the least (p=0.008). Using a vacutainer instead of a syringe for blood collection slightly increased hemolysis (p=0.023).

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionMethod of fluid acquisition
    Needle
    Needleless connector
    Syringe draw
    Vacuum tube
    Biospecimen AcquisitionNeedle gauge
    16
    18
    22
    Biospecimen AcquisitionType of collection container/solution
    Syringe
    Vacutainer tube