Different collection tubes in cardiac biomarkers detection.
Daves M, Trevisan D, Cemin R
Purpose
The purpose of this paper was to determine the effects of anticoagulant and gel-separator on detection of cardiac biomarkers in blood specimens from patients with acute coronary syndrome.
Conclusion
Collection tube type and anticoagulant had no effects on cardiac troponin I. Decreased myoglobin and CK-MB were found in specimens collected in heparin tubes without a gel separator compared to specimens collected in gel separator tubes with no anticoagulant.
Studies (1)
Purpose
The purpose of this study was to determine the effects of heparin and gel separator tubes on the detection of cardiac troponin I, myoglobin ad CK-MB in blood specimens from patients with acute coronary syndrome.
Summary of Findings
Collection tube type and anticoagulant had no effects on cardiac troponin I. Decreased myoglobin and CK-MB were found in specimens collected in heparin tubes without a gel separator compared to specimens collected in gel separator tubes with no anticoagulant.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Coronary Artery DiseasePlatforms
ProteinImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant HeparinNoneBiospecimen Acquisition Type of collection container/solution Gel separatorPlainImmunoassay Specific Targeted peptide/protein CK-MBCardiac troponin IMyoglobin