Evaluation and advantages of an automatic magnetic mixing of syringes integrated to a whole blood gas analyser.

Benoit MO, Paul JL

Publication: Scand J Clin Lab Invest (2009) • Vol 69 • Page 628-32PMID: 19391061View on PubMed

Purpose

The purpose of this paper was to evaluate the effect of automatic mixing using the ABL 825 FLEX on total hemoglobin (tHb) and potassium (K+) in blood specimens.

Conclusion

No significant differences in tHb or K+ were found between specimens analyzed using the ABL 825 FLEX and those analyzed on the LH 750 (tHb) and the LX20 (K+).

Studies (1)

Purpose

The purpose of this study was to determine the effects of magnetic versus turnaround mixing on the determination of tHb and K+. Specimens were collected in SafePICO syringes and placed into the ABL 825 FLEX machine for magnetic mixing and analysis of tHb and K+, collected in an EDTA tube and mixed by turnaround mixing and analyzed for tHb on a LH750, or collected in heparinized tubes, centrifuged and analyzed for K+ on a LX20.

Summary of Findings

No significant differences in tHb or K+ were found between specimens mixed and analyzed using the ABL 825 FLEX and those mixed by turn around mixing and analyzed on the LH 750 (tHb) and the LX20 (K+). However, a slight, clinically insignificant, positive bias was found for specimens analyzed on the ABL 825 FLEX machine. The authors attribute the bias to differences in the measurement methods used by the two autoanalyzers and not to differences in mixing. The authors conclude that the ABL 825 FLEX can be used for analysis of tHb and K+ in blood specimens.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinHematology/ auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Clinical chemistry/auto analyzer SpecificTechnology platform
    ABL 825 FLEX
    LH750
    LX20