Stability of PO2, PCO2, and pH in fresh blood samples stored in a plastic syringe with low heparin in relation to various blood-gas and hematological parameters.

Beaulieu M, Lapointe Y, Vinet B

Publication: Clin Biochem (1999) • Vol 32 • Page 101-7PMID: 10211625View on PubMed

Purpose

The purpose of this paper was to determine the stability of partial pressure oxygen (PO2), partial pressure carbon dioxide (pCO2) and pH in blood specimens and the effects of storage on hematological parameters.

Conclusion

PO2 decreased with time when specimens were stored at 22 degrees C, but increased with time when stored on ice. However, a large degree of variability in PO2 was observed between specimens stored at either temperature. The change in PO2 did not correlate with leukocyte count, platelet count or hemoglobin concentration, but in specimens stored at 22 degrees C, the change in PO2 did correlate with the oxygen saturation level of hemoglobin (SatHbO2). PCO2 increased with storage at either temperature, regardless of initial levels, while pH decreased. The magnitude of the changes in PO2, PCO2, and pH were much greater in specimens stored at 22 degrees C than in those stored on ice.

Studies (1)

Purpose

The purpose of this study was to determine the stability of PO2, pCO2 and pH in blood specimens and the effects of storage on hematological parameters.

Summary of Findings

PO2 decreased with time when specimens were stored at 22 degrees C (p<0.001), but increased with time when stored on ice (p<0.005). However, a large degree of variability in PO2 was observed between specimens stored at either temperature, and in specimens with an initial PO2 of less than 50 mm mercury, the changes with storage were insignificant. The change in PO2 did not correlate with leukocyte count, platelet count or hemoglobin concentration, but in specimens stored at 22 degrees C, the change in PO2 did correlate with SatHbO2. PCO2 increased with storage at either temperature, regardless of initial levels, while pH decreased. The magnitude of the changes in PO2, PCO2, and pH were much greater in specimens stored at 22 degrees C than in those stored on ice.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Small moleculepH
    Cell count/volumeHematology/ auto analyzer
    ProteinSpectrophotometry
    GasRadiometer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Heparin
    StorageStorage duration
    0 min
    120 min
    15 min
    30 min
    60 min
    StorageStorage temperature
    22 degrees C
    On ice
    StorageType of storage container
    Plastic syringe