Skip to main content

Stability of blood gases, electrolytes and haemoglobin in heparinized whole blood samples: influence of the type of syringe.

Müller-Plathe O, Heyduck S

Publication: Eur J Clin Chem Clin Biochem (1992) • Vol 30 • Page 349-55PMID: 1511069View on PubMed

Purpose

The purpose of this paper was to compare the stability of blood gas, electrolytes and hemoglobin stored in 5 different brands of polypropylene syringes or in glass syringes.

Conclusion

After 45 minutes of storage in ice water, specimens in plastic syringes generally had increased partial pressure oxygen (pO2), and partial pressure carbon dioxide (pCO2), and a trend toward decreased pH. Slight changes in sodium, calcium, potassium and hemoglobin were also observed in specimens stored for 45 min in ice water.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage and syringe type on blood gas, electrolytes and hemoglobin. 5 commercial polypropylene syringes with different fill volumes, concentrations and types of heparin, and electrolyte balancing were compared to fluid electrolyte balanced 5 mL glass syringes with 10 IU/10^3 L blood of sodium heparin salt.

Summary of Findings

All plastic syringes showed an increase in pO2 after storage for 45 minutes in ice water. The increase was least in hypoxemic specimens or when the Braun Inject (B2) syringes were used. The increase was greatest in plasma compared to whole blood or when Braun Omnifix (B1) syringes were used. The increase was even greater and apparent after 10 minutes instead of 20 min when the blood was stored at room temperature in B1 syringes instead of glass syringes. The pCO2 in specimens in plastic syringes increased slightly with storage in ice water, except when stored in the B1 syringe or when hyperkapnic specimens were used. A slight trend toward decreasing pH after 45 minutes of storage in ice water was noted, but the change was generally below 0.01. Slight changes in sodium, calcium, and potassium were observed, but the amounts were dependent on heparin and syringe type. The changes in hemoglobin were small, but the reproducibility was poor, especially for specimens stored in the B1 syringe.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Small moleculepH
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    Sodium heparin
    Biospecimen AcquisitionType of collection container/solution
    Braun Inject
    Braun Omnifix
    Glass syringe
    Radiometer Art Samp B129
    Starstedt monvette LH
    Terumo Preza-pak II
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Whole blood
    StorageStorage duration
    0 min
    10 min
    20 min
    45 min
    StorageStorage temperature
    Ice water
    Room temperature