Influence of syringe material on arterial blood gas determinations
Winkler JB, Huntington CG, Wells DE, Befeler B
Publication: Chest (1974) • Vol 66 • Page 518
Purpose
The purpose of this paper was to determine the effects of syringe material and storage at room temperature on the measurement of blood gas.
Conclusion
The differences between polypropylene and glass containers were within the accepted errors of the method and no consistent effects were seen for any analyte. Further, no effects of storage at room temperature for up to 150 min were found on any analyte.
Studies (1)
Purpose
The purpose of this study was to determine the effects of syringe material and storage at room temperature on the measurement of blood gas.
Summary of Findings
The differences between polypropylene and glass containers were within the accepted errors of the method and no consistent effects were seen on the measured pH, partial pressure carbon dioxide, base excess, partial pressure oxygen, or oxygen saturation. Further, no effects of storage at room temperature for up to 150 min were found on any analyte.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculepHGasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant HeparinBiospecimen Acquisition Type of collection container/solution Glass syringePolypropylene syringeStorage Time at room temperature 0 min120 min150 min30 min60 min90 min