Lipid emulsion solution: A novel cause of hemolysis in serum and plasma blood samples.

Jaben EA, Koch CD, Karon BS

Publication: Clin Biochem (2011) • Vol 44 • Page 254-6PMID: 21070759View on PubMed

Purpose

The purpose of this paper was to determine the effects of lipid emulsion solution on hemolysis of blood specimens.

Conclusion

Hemolysis was proportional to the amount of lipid emulsion solution relative to total parenteral nutrition (TPN) regardless of the tube type, but only increased with time at room temperature when an equal amount of lipid emulsion solution and TPN were added and a plasma separator tube was used. Specimens in plasma separator tubes had significantly more hemolysis than those in blood gas syringes at concentrations of lipid emulsion solution and TPN greater than 10%.

Studies (1)

Purpose

The purpose of this study was to investigate the effects of container type, ratio and concentration of lipid emulsion solution and time at room temperature on hemolysis of blood specimens. Blood samples were spiked with different ratios of lipid emulsion solution to TPN before being aliquoted into plasma separator tubes or blood gas syringes.

Summary of Findings

When an equal amount of lipid emulsion solution and TPN were added to the blood specimen at a final concentration of 10%, gross hemolysis (hemoglobin >300 mg/dL), which increased with time at room temperature, was observed in the plasma separator tube specimens. In contrast, moderate hemolysis (150 mg/dL), which was not dependent on time at room temperature, was observed in the blood gas syringe specimens when an equal amount of lipid emulsion solution and TPN were added at a final concentration of 10%. As the ratio of lipid emulsion solution to TPN decreased, the hemolysis became less pronounced, regardless of container, and the effect of time at room temperature disappeared. Only minimal hemolysis was observed when equal amounts of lipid emulsion solution and TPN were added at a concentration of 1%. However at higher concentrations, significant hemolysis was observed in the plasma separator tube specimens, and minimal hemolysis was observed in the blood gas syringe specimens.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Blood gas syringe
    Plasma separator tubes
    Biospecimen Aliquots and ComponentsBiospecimen components
    0% lipid emulsion solution
    1% of 1:1 lipid emulsion solution to TPN
    10% of 1:1 lipid emulsion solution to TPN
    10% of 1:4 lipid emulsion solution to TPN
    10% of 1:8 lipid emulsion solution to TPN
    15% of 1:1 lipid emulsion solution to TPN
    20% of 1:1 lipid emulsion solution to TPN
    5% of 1:1 lipid emulsion solution to TPN
    StorageTime at room temperature
    0 min
    10 min
    20 min
    30 min
    60 min