Thrombin generation is extremely sensitive to preheating conditions.
De Smedt E, Hemker HC
Purpose
The purpose of this paper was to investigate the effects of preheating time on thrombin generation in plasma.
Conclusion
The endogenous thrombin potential was overestimated by 50% when specimens were preheated in the instrument for 2 min. Preheating in the instrument for more than 2 minutes resulted in specimen evaporation and up to a 4 degree C difference in temperature between wells. When specimens were preheated for 5-15 min on the plate heater, thrombin generation became consistent.
Studies (1)
Purpose
The purpose of this study was to investigate the effects of preheating time on thrombin generation in plasma.
Summary of Findings
The endogenous thrombin potential was overestimated by 50% when specimens were preheated in the instrument for 2 min. Preheating in the instrument for more than 2 minutes resulted in specimen evaporation and up to a 4 degree C difference in temperature between wells. When specimens were preheated for 5-15 min on the plate heater, thrombin generation became consistent.
Biospecimens
FluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 0.5 pM tissue factor5 pM tissue factorHematology/ auto analyzer Specific Heating method Covered on plate-heaterIn fluorometerHematology/ auto analyzer Specific Incubation temperature 37 degrees CHematology/ auto analyzer Specific Incubation time 10 min15 min2 min5 min