The purpose of this paper was to determine the effects of collection method, storage temperature, storage duration, and freeze-thaw cycling on protein profiles of urine specimens.
Sample handling for mass spectrometric proteomic investigations of human urine
Petri AL, Høgdall CK, Christensen IJ, Simonsen AH, T'Jampens D, Hellmann ML, Kjaer SK, Fung E T, Høgdall EV
Purpose
Conclusion
Use of the CM10 array resulted in the identification of 122 protein peaks, of which 28 were unstable. In contrast, 65 peaks were detected using the IMAC30 array, only one of which was unstable. Of the 29 unstable peaks identified using either array, 11 were significantly affected by urine type (first morning urine, midday urine, or catheterized urine) 5 were significantly affected by storage temperature, and 19 were significantly affected by storage duration. The effect of storage duration was greater when comparing 1 hour or 6 hours with 72 hours of storage, rather than comparing 1 and 6 hours directly. Freeze-thaw cycling did not significantly affect any protein peaks.
Studies (1)
Purpose
Summary of Findings
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
NeoplasticBenignNeoplasticCarcinomaPlatforms
PeptideSELDI-TOF MSProteinSELDI-TOF MSProteinColorimetric assaySmall moleculeColorimetric assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Patient gender FemaleBiospecimen Acquisition Method of fluid acquisition Catheterized urineVoided urine (24-h collection)Biospecimen Acquisition Time of biospecimen collection First morning urineMidday urineStorage Freeze/thaw cycling 1 cycle2 cycles3 cycles4 cycles5 cycles6 cycles7 cycles8 cyclesStorage Storage duration 1 h6 h72 hStorage Storage temperature 20 degrees C4 degrees CSELDI-TOF MS Specific Type of array CM10IMAC30