Validation of an automated intact N-terminal propeptide of type I procollagen (PINP) assay.

Koivula MK, Richardson J, Leino A, Valleala H, Griffiths K, Barnes A, Konttinen YT, Garrity M, Risteli J

Publication: Clin Biochem (2010) • Vol 43 • Page 1453-7PMID: 20923676View on PubMed

Purpose

The purpose of this paper was to validate an automated assay for N-terminal propeptide of type 1 procollagen (PINP) in serum specimens.

Conclusion

When serum was stored at room temperature or 4 degrees C for up to 168 h, no change in intact or total PINP was found, but when serum was stored at 37 degrees C, both intact and total PINP began to decrease after 8 h of storage. A correlation of 0.986 was found for intact PINP measured by radioimmunoassay (RIA) and autoanalyzer in hemodialysis patients. No linear correlation was found between intact and total PINP measured by autoanalyzer in hemodialysis patients, but correlations of 0.947 before treatment and 0.966 after treatment were found for rheumatoid arthritis patients.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage, assay type, and disease on intact and total PINP in serum specimens from control, renal failure and rheumatoid arthritis patients. All patient specimens were stored at -20 degrees C.

Summary of Findings

When serum was stored at room temperature or 4 degrees C, for up to 168 h, no change in intact or total PINP was found, but when stored at 37 degrees C both intact and total PINP began to decrease after 8 h of storage. A correlation of 0.986 was found for intact PINP measured by RIA and autoanalyzer in hemodialysis patients. No linear correlation was found between intact and total PINP measured by autoanalyzer in hemodialysis patients, but correlations of 0.947 before treatment and 0.966 after treatment were found for rheumatoid arthritis patients.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Other diagnoses
    Normal
    Rheumatoid Arthritis
  • Platforms

    PeptideRadioimmunoassay
    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Renal failure
    PreaquisitionOther drugs
    Bisphosphonates
    Corticosteroids
    Disease modifying antirheumatic drugs
    StorageStorage duration
    0 h
    120 h
    144 h
    168 h
    24 h
    4 h
    48 h
    72 h
    8 h
    96 h
    StorageStorage temperature
    37 degrees C
    4 degrees C
    Room temperature