Saliva analysis by surface-enhanced laser desorption/ionization time-of-flight mass spectrometry (SELDI-TOF/MS): from sample collection to data analysis.

Papale M, Pedicillo MC, Di Paolo S, Thatcher BJ, Lo Muzio L, Bufo P, Rocchetti MT, Centra M, Ranieri E, Gesualdo L

Publication: Clin Chem Lab Med (2008) • Vol 46 • Page 89-99PMID: 18020972View on PubMedSuggested by: ISBER

Purpose

The purpose of this paper was to determine the effects of different specimen collection, handling, storage, and analysis methods on protein profiles of saliva.

Conclusion

In the absence of protease inhibitors, saliva protein profiles showed reduced peak intensities after 3 hours at room temperature. However, the addition of protease inhibitor cocktail to the specimens eliminated this effect. Up to 5 freeze-thaw cycles had little effect on saliva protein profiles. Proteomic profiles of individuals were found to change significantly throughout the day. Significant differences based on gender were found with each chip array type. When one individual specimen was analyzed on all four chip arrays, only 14-25% of the peaks were shared between all four array types. Centrifugation of saliva specimens prior to analysis did not alter the number of peaks observed. Removal of glycoproteins by ethanol precipitation increased the number of peaks and improved the intensity and resolution. Denaturation increased the number of peaks for all arrays except the strong cationic exchange CM10 chip. The number of peaks also changed based on detection by alpha-cyano-hydroxycinnamic acid (CHCA) or sinapinic acid (SA).

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage duration, freeze-thaw cycling, collection time, denaturing agents, depletion of glycoproteins, centrifugation, matrix type, array type, gender, and individual on protein profiles of saliva. Specimens were stored at -80 degrees C until analyzed.

Summary of Findings

In the absence of protease inhibitors, saliva protein profiles showed reduced peak intensities after 3 hours at room temperature indicating specimen degradation. However, the addition of protease inhibitor cocktail to the specimens resulted in similar peak intensities between 0 and 3 hours at room temperature. Up to 5 freeze-thaw cycles had little effect on saliva protein profiles, regardless of the addition of protease inhibitors. Proteomic profiles of individuals were found to change significantly throughout the day, which the authors believe may be partially attributable to food ingestion. Significant differences based on gender were found with each chip array type. When one individual specimen was analyzed on all four chip arrays, only 14-25% of the peaks were shared between all four array types. Centrifugation of saliva specimens at 1300 x g prior to analysis did not alter the number of peaks observed. Removal of glycoproteins by ethanol precipitation increased the number of peaks and improved the intensity and resolution. Denaturation increased the number of peaks for all arrays except the strong cationic exchange CM10 chip. The highest number of peak was obtained with different buffers for different arrays. Further, the number of peaks also changed based on detection by CHCA or SA.

  • Biospecimens

    FluidSaliva
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    PeptideSELDI-TOF MS
    ProteinSELDI-TOF MS
    ProteinColorimetric assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient gender
    Female
    Male
    Biospecimen AcquisitionTime of biospecimen collection
    After lunch
    Before bed
    Before lunch
    Upon waking
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifuged
    Not centrifuged
    StorageFreeze/thaw cycling
    1 cycle
    2 cycles
    3 cycles
    4 cycles
    5 cycles
    StorageTime at room temperature
    0 h
    3 h
    4 h
    48 h
    Analyte Extraction and PurificationProtease inhibitor
    Cocktail
    No protease inhibitor added
    SELDI-TOF MS SpecificDenaturation
    DTT
    Urea and CHAPS
    Urea, CHAPS and DTT
    SELDI-TOF MS SpecificDetection method
    CHCA
    SPA
    SELDI-TOF MS SpecificType of array
    Hydrophobic interaction
    Metal affinity
    Strong anionic exchange
    Strong cationic exchange