Performance characteristics of six intact parathyroid hormone assays.
La'ulu SL, Roberts WL
Purpose
The purpose of this paper was to determine the effects of collection tube type, storage and assay on parathyroid levels in blood specimens.
Conclusion
Compared to specimens collected in plain red top tubes, significant effects of collection tube type were found when specimens were collected in plasma separator (PST), lithium heparin or EDTA tubes and analyzed with the IMMULITE assay or collected in EDTA and analyzed with the E170 assay. The majority of specimens stored frozen for 3 days or refrigerated for 2 days were significantly different from fresh specimens for all immunoassay platforms tested, but only 6 combinations of tube type, storage temperature and duration and assay type were found to be clinically significant when a cutoff of changes greater than 12.5% was applied.
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type, storage and assay on parathyroid levels in blood specimens.
Summary of Findings
Compared to specimens collected in plain red top tubes, significant effects of collection tube type were found when specimens were collected in PST, lithium heparin or EDTA tubes and analyzed with the IMMULITE assay or collected in EDTA and analyzed with the E170 assay. The majority of specimens stored frozen for 3 days or refrigerated for 2 days were significantly different from fresh specimens for all immunoassay platforms tested, but only 6 combinations of tube type, storage temperature and duration and assay type were found to be clinically significant when a cutoff of changes greater than 12.5% was applied. No clinically significant changes were found in specimens stored refrigerated for 24 h when compared to fresh specimens. Specimens refrigerated for 48 h were clinically different from fresh specimens when collected in a serum separator tube (SST) and analyzed with an ACCESS or ADVIA Centaur assay or when collected in a PST tube and analyzed using the IMMULITE 2000 assay. Specimens stored frozen for 3 days were clinically different from fresh specimens when collected in a PST tube and analyzed using an IMMULITE 2000 assay or when collected in a lithium heparin tube and analyzed using the ADVIA Centaur or LIAISON assays.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
PeptideImmunoassayPeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTALithium heparinBiospecimen Acquisition Type of collection container/solution Plain serumPlasma separator tube (PST)Serum separator tube (SST)Storage Storage duration 0 days1 day2 days3 daysStorage Storage temperature -70 degrees C2-8 degrees CClinical chemistry/auto analyzer Specific Technology platform ACCESS 2 (Beckman)ADVIA Centaur (Siemens)ARCHITECT i2000 (Abbott)IMMULITE 2000 (Siemens)LIAISON (DiaSorin)Modular E170 (Roche)