Quality assurance in analysis of cryogenically stored liver tissue specimens from the NIST National Biomonitoring Specimen Bank (NBSB).
Mackey EA, Demiralp R, Fitzpatrick KA, Porter BJ, Wise SA, Becker PR, Greenberg RR
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine if the temperature (-80 or -150 degrees C) of long term specimen storage affects trace element composition. Subsets of five liver specimens snap-frozen and stored in pre-cleaned Teflon jars at -150 degrees in a liquid nitrogen vapor cooled freezer or a -80 degree C freezer for 17 years were examined.
Summary of Findings
No macroscopic evidence of specimen degradation was observed among cryopreserved liver specimens stored at either temperature. Specimen trace element composition (Rb, Fe, Zn) was unaltered by 17 years of storage at either storage temperature. The following elements were not altered by prolonged storage at either temperature, although several elements were at or below the detection limit of the platform: As, Mo, Cd, Sm, La.
Biospecimens
TissueLiverPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
Electrolyte/MetalAtomic absorption spectroscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 17 y7 yStorage Storage temperature -150 degrees C-80 degrees C