A model for an uncertainty budget for preanalytical variables in clinical chemistry analyses.

Rynning M, Wentzel-Larsen T, Bolann BJ

Publication: Clin Chem (2007) • Vol 53 • Page 1343-8PMID: 17495014View on PubMedSuggested by: ISBER

Purpose

The purpose of this paper was to determine the effects of preanalytical variables on clinical chemistry analytes in blood specimens

Conclusion

The use of plain glass tubes versus SST tubes significantly decreased serum levels of creatinine, but no effect of using glass versus plastic SST tubes was found. Increasing the clotting time from 40-70 minutes to 120-150 minutes significantly decreased glucose and calcium. Storing serum for 48 h at room temperature, prior to freezing at -80 degrees C, significantly increased creatinine and decreased glucose. Significantly lower levels of creatinine and higher levels of glucose were measured by a Modular II instrument when compared to levels measured by a Modular I instrument. Centrifugation speed had no effect on any of the analytes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type, clotting time, centrifugation speed (1300 x g versus 2350 x g), storage of serum, and instrument type on serum glucose, calcium, magnesium and creatinine levels.

Summary of Findings

The use of plain glass tubes versus SST tubes significantly decreased serum levels of creatinine (p=0.003), but no effect of using glass versus plastic SST tubes was found (p>0.07). Increasing the clotting time from 40-70 minutes to 120-150 minutes, significantly decreased glucose (p<0.001) and calcium (p<0.001). Storing serum for 48 h at room temperature, prior to freezing at -80 degrees C, significantly increased creatinine (p<0.001) and decreased glucose (p<0.001) when compared to serum left at room temperature for less than 4 hours before freezing. Significantly lower levels of creatinine (p=0.01) and higher levels of glucose (p<0.007) were measured by a Modular II instrument when compared to levels measured by a Modular I instrument. Centrifugation speed had no effect on any of the analytes.

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass serum separator (SST) (Becton Dickinson)
    Plastic serum separator (SSTII) (Becton Dickinson)
    Vacutainer (Becton Dickinson)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    Multiple speeds compared
    StorageStorage duration
    120-150 min
    40-70 min
    StorageStorage temperature
    -80 degrees C
    StorageTime at room temperature
    48 h
    < 4 h
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Roche Modular I
    Roche Modular II