The purpose of this paper was to determine the effects of preanalytical variables on clinical chemistry analytes in blood specimens
A model for an uncertainty budget for preanalytical variables in clinical chemistry analyses.
Rynning M, Wentzel-Larsen T, Bolann BJ
Purpose
Conclusion
The use of plain glass tubes versus SST tubes significantly decreased serum levels of creatinine, but no effect of using glass versus plastic SST tubes was found. Increasing the clotting time from 40-70 minutes to 120-150 minutes significantly decreased glucose and calcium. Storing serum for 48 h at room temperature, prior to freezing at -80 degrees C, significantly increased creatinine and decreased glucose. Significantly lower levels of creatinine and higher levels of glucose were measured by a Modular II instrument when compared to levels measured by a Modular I instrument. Centrifugation speed had no effect on any of the analytes.
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type, clotting time, centrifugation speed (1300 x g versus 2350 x g), storage of serum, and instrument type on serum glucose, calcium, magnesium and creatinine levels.
Summary of Findings
The use of plain glass tubes versus SST tubes significantly decreased serum levels of creatinine (p=0.003), but no effect of using glass versus plastic SST tubes was found (p>0.07). Increasing the clotting time from 40-70 minutes to 120-150 minutes, significantly decreased glucose (p<0.001) and calcium (p<0.001). Storing serum for 48 h at room temperature, prior to freezing at -80 degrees C, significantly increased creatinine (p<0.001) and decreased glucose (p<0.001) when compared to serum left at room temperature for less than 4 hours before freezing. Significantly lower levels of creatinine (p=0.01) and higher levels of glucose (p<0.007) were measured by a Modular II instrument when compared to levels measured by a Modular I instrument. Centrifugation speed had no effect on any of the analytes.
Biospecimens
FluidBloodFluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass serum separator (SST) (Becton Dickinson)Plastic serum separator (SSTII) (Becton Dickinson)Vacutainer (Becton Dickinson)Biospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedMultiple speeds comparedStorage Storage duration 120-150 min40-70 minStorage Storage temperature -80 degrees CStorage Time at room temperature 48 h< 4 hClinical chemistry/auto analyzer Specific Technology platform Roche Modular IRoche Modular II