UK Biobank Pilot Study: stability of haematological and clinical chemistry analytes.

Jackson C, Best N, Elliott P

Publication: Int J Epidemiol (2008) • Vol 37 Suppl 1 • Page i16-22PMID: 18381388View on PubMedSuggested by: ISBER

Purpose

The purpose of this paper was to assess the effects of storage on clinical chemistry and hematological analytes in plasma, serum and urine specimens.

Conclusion

Of the 46 analytes measured, 23 changed by less than 1% per 12 h of storage and another 21 changed between 1 and 3% per 12 h of storage. The largest changes were observed for insulin, which increased by 3.9% per 12 h of storage, and eosinophils, which decreased by 12% per 12 h of storage. However, both changes were highly variable between individuals and not statistically significant. The authors conclude that changes over the first 36 h of storage are small by comparison to individual variability and assay error.

Studies (1)

Purpose

The purpose of this study was to assess the effects of storage on clinical chemistry and hematological analytes in plasma, serum and urine specimens.

Summary of Findings

Of the 46 analytes measured, 23 changed by less than 1% per 12 h of storage and another 21 changed between 1 and 3% per 12 h of storage. The largest changes were observed for insulin, which increased by 3.9% per 12 h of storage, and eosinophils, which decreased by 12% per 12 h of storage. However, both changes were highly variable between individuals and not statistically significant. Using a probability model, calcium, cholesterol, fibrinogen and HDL cholesterol all had a probability of a negative bias over time between 0.9 and 0.975. None of the analytes showed a high probability of a positive bias over time. More than 72% of the variability in levels was due to baseline values, and much of the remaining variability was attributable to measurement error. The authors conclude that changes over the first 36 h of storage are small by comparison to individual variability and assay error.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidUrine
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    0 h
    12 h
    24 h
    36 h
    StorageStorage temperature
    -196 degrees C
    -80 degrees C