Preanalytical impact of sample handling on proteome profiling experiments with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry.

Findeisen P, Sismanidis D, Riedl M, Costina V, Neumaier M

Publication: Clin Chem (2005) • Vol 51 • Page 2409-11PMID: 16306114View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage of serum and plasma specimens at room temperature on proteome profiles.

Conclusion

Major changes were observed in the proteome profiles of serum when specimens were stored at room temperature for 4 h after centrifugation but before freezing at -80 degrees C, regardless of the resins used for sample processing. The authors conclude that plasma is less sensitive to preanalytical handling and may be more suitable for proteomic analysis than serum, however, the time-dependant degradation of a 1467 Dalton peak may be an appropriate quality indicator for inclusion/exclusion of serum specimens.

Studies (1)

Purpose

The purpose of this study was to determine the effects of anticoagulant, type of resin used for sample processing, and storage of serum and plasma at room temperature after centrifugation on proteome profiles.

Summary of Findings

Compared to specimens processed and frozen immediately at -80 degrees C, major changes were observed in the proteome profiles of serum when specimens were stored at room temperature for 4 h before freezing, regardless of the resins used for sample processing. After 4 h at room temperature, some changes were also seen in heparin-plasma profiles with weak cation-exchange chromatography and in EDTA-plasma profiles with hydrophobic interaction or immobilized metal-ion affinity chromatography. These changes were not as profound as those seen in serum. A single 1467 Dalton peak, identified as part of the N-terminal portion of fibrinogen alpha, was found to decrease in intensity with extended storage at room temperature. The authors conclude that plasma is less sensitive to preanalytical handling and may be more suitable for proteomic analysis than serum, however, the intensity of the 1467 Dalton peak may be appropriate as a quality indicator for inclusion/exclusion of serum specimens.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
    Not specified
  • Platforms

    ProteinMALDI-TOF MS
    PeptideMALDI-TOF MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Ammonium-heparin
    EDTA
    None
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    StorageStorage temperature
    -80 degrees C
    StorageTime at room temperature
    0 h
    1 h
    2 h
    4 h
    MALDI-TOF MS SpecificResin
    Hydrophobic interaction chromatography C8
    Immobilized metal-ion affinity chromatography
    Weak cation-exchange chromatography