The purpose of this paper was to evaluate the effects of partial tube filling and storage on blood gas parameters.
Potential preanalytical errors in whole-blood analysis: effect of syringe sample volume on blood gas, electrolyte and lactate values.
Hedberg P, Majava A, Kiviluoma K, Ohtonen P
Purpose
Conclusion
Filling the 3.0 mL tube to 1.5 or 1.8 mL resulted in changes of less than 10% in oxygen partial pressure (pO2), oxygen saturation (O2 sat), and sodium from both arterial and venous collections and a 14.8% decrease in carboxyhemoglobin in total hemoglobin (COHb) in 1.5 mL collections of arterial, but not venous blood. The use of 1.0 mL fill volume led to significant changes in most variables; however, only pO2, base excess in blood (BEvT) and CoHb were changed by more than 10% in venous and/or arterial blood specimens. Storage of 1 mL blood specimens at 4 or 23 degrees C for up to 90 minutes resulted in changes of more than 10% in partial pressure of carbon dioxide (pCO2), lactate, COHb, BEvt and methemoglobin in total hemoglobin (MetHb).
Studies (2)
Purpose
The purpose of this study was to evaluate the effects of partial tube filling on blood gas parameters.
Summary of Findings
Filling the 3.0 mL tube to 1.8 mL resulted in significant changes of less than 10% in pO2, O2 sat, and sodium from both arterial and venous collections and in pH from arterial and calcium from venous collections. Using 1.5 mL caused similar changes of less than 10% in pO2, O2 sat, sodium and lactate in both arterial and venous specimens and a 14.8% decrease in COHB in arterial but not venous blood. The use of 1.0 mL fill volume led to significant changes in most variables, but only pO2, BEvT and CoHb were changed by more than 10% in venous and/or arterial blood specimens.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
GasClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinHematology/ auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anatomical location of blood draw ArteryVeinBiospecimen Aliquots and Components Aliquot size/volume 1.0 mL in 3.0 mL tube1.5 mL in 3.0 mL tube1.8 mL in 3.0 mL tube3.0 mL fill volume
Purpose
Summary of Findings
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerProteinHematology/ auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume 1.0 mL in 3.0 mL tubeStorage Storage duration 1-5 min15 min30 min60 min90 minStorage Storage temperature 23 degrees C4 degrees C