Methods for extracting and amplifying genomic DNA isolated from frozen serum.
Dixon SC, Horti J, Guo Y, Reed E, Figg WD
Purpose
The purpose of this paper was to compare the yield and quality of DNA obtained from serum using different DNA extraction kits.
Conclusion
DNA extraction using the nonorganic extraction kit, Centrifree, GlassMax or by proteinase K digestion followed by centrifugation in Ultrafree Probind or MicroCon-100 columns resulted in DNA that was either protein contaminated or too dilute. All other methods resulted in DNA of sufficient concentration and quality as determined by spectrophotometer. When QIAamp Blood kit, Chelex-100, Boiling and Easy DNA methods were further examined by PCR amplification, it was found that DNA extracted with QIAamp Blood and EasyDNA kits was able to be amplified from only 55-70% of specimens. The rate of amplification success could be increased to 88.5% if DNA that did not amplify when isolated with QIAamp was then isolated with EasyDNA. In addition, the use of SYBR-green instead of ethidium bromide to visualize PCR products led to increased sensitivity and decreased background.
Studies (2)
Purpose
The purpose of this study was to compare the yield and quality of DNA extracted from serum using 13 different DNA extraction kits and protocols.
Summary of Findings
Protein contamination was a problem when DNA was extracted by the nonorganic extraction kit or by proteinase K digestion followed by centrifugation in Ultrafree Probind or MicroCon-100 columns. DNA extracted with Centrifree or GlassMax was of good quality, but too dilute for PCR without concentration. All other methods produced DNA of similar quality and sufficient concentration as determined by spectrophotometer.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
NormalPlatforms
DNASpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method BoilingCentrifree units (Amicon)DNAzol (Life technologies)Easy-DNA kit (Invitrogen)Extraction with Chelex-100 (BioRad)GlassMax DNA isolation spin cartridge system (Life technologies)Nonorganic DNA extraction kit (Oncor)Proteinase K then MicroCon-100 columns (Amicon)Proteinase K then Ultrafree Probind columns (Millipore)Proteinase K with Phenol ChloroformQIAamp Blood kit (Qiagen)ReadyAmp genomic DNA purification kit (Promega)Split Second DNA preparation kit (Boehringer Mannheim)
Purpose
The purpose of this study was to compare the yield, quality and amplificability of DNA extracted from serum using 4 different methods. The effect of visualizing electrophoresed PCR products with ethidium bromide versus SYBR green I was also examined.
Summary of Findings
DNA yield from the four methods was highly variable with the highest average yield obtained when specimens were extracted with Chelex-100 and the lowest with heat extraction. However, both Chelex-100 and heat extraction resulted in DNA for which amplification of the trinucleotide region of the androgen receptor was successful in only 10-20% of specimens. In contrast, the same region was successfully amplified 55-70% of the time when DNA was isolated with either QIAamp or Easy-DNA. For 18.2% of specimens, amplification was successful when QIAamp, but not EasyDNA was used, and for 20.9%, amplification was successful when EasyDNA, but not QIAamp was used. 15.5% of specimens failed to amplify with either method. The use of SYBR-green to visualize DNA products led to increased sensitivity and decreased background resulting in only 11.4% of specimens unamplifiable by either method.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
NormalPlatforms
DNASpectrophotometryDNAPCRDNAElectrophoresisPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Analyte isolation method BoilingChelex-100 (BioRad)Easy-DNA kit(Invitrogen)QIAamp Blood kit(Qiagen)Electrophoresis Specific Detection method Ethidium bromideSYBR Green I