Immunoassay interference by a commonly used blood collection tube additive, the organosilicone surfactant silwet L-720.
Bowen RA, Chan Y, Ruddel ME, Hortin GL, Csako G, Demosky SJ Jr, Remaley AT
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine if the surfactant Silwet L-720 interferes with serum measurement of clinical chemistry analytes by competitive immunoassays. All serum was processed and analyzed within 3 h at RT, or 3 days at 4 degrees C.
Summary of Findings
The apparent concentrations of TT3, progesterone, cortisol, TT4 and TBG as measured by competitive immunoassays on an Immulite but not AxSym analyzer were all increased with increasing concentrations of Silwet L-720; however the concentrations of SHBG and TSH, which were measured by non-competitive assays, were not affected. Specifically, the surfactant was found to desorb antibodies from the bead surface.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
PeptideClinical chemistry/auto analyzerGlycoproteinClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution 0-400 mg/L Silwet L-720 surfactant to mimic SST tubesClinical chemistry/auto analyzer Specific Technology platform AxSYMImmulite 2500