Effect of blood collection tubes on total triiodothyronine and other laboratory assays.

Bowen RA, Chan Y, Cohen J, Rehak NN, Hortin GL, Csako G, Remaley AT

Publication: Clin Chem (2005) • Vol 51 • Page 424-33PMID: 15576427View on PubMed

Purpose

The purpose of this paper was to determine the effects of blood collection tube type on 17 immunoassays, 30 clinical chemistry tests and 33 immunology assays in serum.

Conclusion

Collection of blood into SST tubes instead of glass tubes resulted in significant differences in 16 analytes. Of the observed changes only the increased total triiodothyronine (TT3) in serum from SST tubes compared with glass tubes was clinically significant. The effect of tube type only occurred when specimens were analyzed on an Immulite analyzer and was dependent on tube lot number but not specimen dilution or type of clot activator. The authors to believe the silicone lubricant and coating on the SST tubes may be responsible.

Studies (1)

Purpose

The purpose of this study was to determine the effects of blood collection tube type on 17 immunoassays, 30 clinical chemistry tests and 33 immunology assays on serum. The effects of serum dilution and analyzer type were also examined. Serum specimens were analyzed within 3 h or stored at 4 degrees C for up to 2 days.

Summary of Findings

Collection of blood into SST tubes instead of glass tubes resulted in increased progesterone (38.6%), TT3 (30.7%), cortisol (14%), thyroid-binding globulin (TBG) (12.5%), total thyroxine (TT4) (12%), insulin (11.9%), Beta 2-microglobulin (B2M) (7.6%), dihydroepiandrostendione sulfate (DHEA-SO4) (6.9%) ferritin (2.9%), creatinine (3.66%), albumin (1%), total calcium (0.43%), and total protein (1.71%). Decreases were seen in the levels of testosterone (-1.6%), serum osmality (-0.68%) and magnesium (-2.27%). However, when blood was collected in vacuette tubes instead of glass only TBG (11.6%), insulin (9.6%), B2M (6.8%) and DHEA-SO4 (6.9%) were significantly altered. None of the 33 tested immunological variables were affected by tube type. Of the differences observed between tube types, the authors state only the increased TT3 in serum from SST tubes compared with glass tubes was clinically significant. Interestingly, the effect of tube type only occurred when specimens were analyzed on an Immulite analyzer and was dependent on tube lot number. Dilution of the specimen and change in clot activator between SST tube types did not remedy the interference. Cleaning the tubes prior to use did alleviate the interference and specimen inversion increased the interference leading the authors to believe the silicone lubricant and coating on the SST tubes may be responsible.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass Becton-Dickinson vacutainer
    Plastic Becton-Dickinson SST vacutainer
    Plastic Greiner Bio-One vacuette
    Biospecimen Aliquots and ComponentsAliquot size/volume
    1:1 dilution
    1:3 dilution
    Clinical chemistry/auto analyzer SpecificTechnology platform
    AxSYM
    Immulite 2000