Skip to main content

Effects of blood-collection systems and tubes on hematologic, chemical, and coagulation tests and on plasma hemoglobin.

Almagor M, Lavid-Levy O

Publication: Clin Chem (2001) • Vol 47 • Page 794-5PMID: 11274046View on PubMed

Purpose

The purpose of this paper was to determine the effects of blood collection systems on hematological and chemical analytes.

Conclusion

Use of the Greiner blood collection system resulted in a nonlinear blood flow and an increase in serum magnesium (Mg), plasma hemoglobin and prothrombin time when compared to the Becton Dickinson (BD) system. The increased serum Mg is believed to be due to optical interference by hemoglobin and the authors suggest measurement of Mg by ion-selective electrode may be necessary. However, despite the hemolysis all test results remained within the normal range.

Studies (1)

Purpose

The purpose of this paper was to compare the use of BD Vacutainer tubes and the Greiner blood collection system when analyzing sodium, potassium, aspartate aminotransferase, alanine aminotransferase amylase, alkaline phosphatase, lactate dehydrogenase, bilirubin, high density lipoprotein (HDL)-cholesterol, magnesium, calcium, differential blood count, activated partial thromboplastin time, prothrombin time, fibrinogen, and hemoglobin.

Summary of Findings

Use of the Greiner blood collection system resulted in a nonlinear blood flow and an increase in serum Mg (2.4%), plasma hemoglobin (31%) and prothrombin time (8%). The increased serum Mg is believed to be due to optical interference by hemoglobin and the authors suggest measurement of Mg by ion-selective electrode may be necessary. However, despite the hemolysis all test results remained within the normal range.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
    ProteinSpectrophotometry
    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Becton Dickinson Vacutainer
    Greiner Vacuette