Measuring cytokine levels in blood. Importance of anticoagulants, processing, and storage conditions.
Thavasu PW, Longhurst S, Joel SP, Slevin ML, Balkwill FR
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to investigate the effects of collection tube, anticoagulant, and storage at 4 degrees C or RT on cytokines in blood and plasma specimens. Specimens were spiked with human recombinant cytokines unless otherwise stated.
Summary of Findings
Cytokine levels in unspiked blood were below the reliable levels of detection, and no evidence of cytokine induction was found. TNF-alpha and IL-6 decreased with time at 4 degrees C (p<0.001) or RT (p<0.001), but the decrease in IL-6 was less pronounced at 4 degrees C. Similarly, IFN-gamma decreased with time at 4 degrees C or RT in EDTA or serum specimens (p<0.0001), but not in heparinized specimens where IFN-gamma recovery was twice as high despite having been similarly spiked. IL-1a and IFN-alpha decreased with time at RT (p<0.001) but not at 4 degrees C. In contrast, IL-1b levels did not change with storage at either temperature in any tube type. No differences in TNF-alpha or IL-1a recovery rates were observed with different anticoagulants or with the addition of protease inhibitors to EDTA, but IL-6 recovery was consistently higher when EDTA was used. TNF-alpha, IL-1a, and IL-6 recovery rates were lower in serum. Use of commercial reagents instead of sterile non-pyrogenic materials gave similar results for all cytokines except IFN-gamma where the effects of time and temperature were lost.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTAHeparinNoneBiospecimen Acquisition Type of collection container/solution Commercial tubesSterile non-pyrogenic material and conditionsBiospecimen Aliquots and Components Biospecimen components Not spikedSpiked with cytokineBiospecimen Aliquots and Components Blood and blood products PlasmaSerumStorage Storage duration 1 h15 min2 h24 h30 min4 hStorage Storage temperature -40 degrees C4 degrees CRoom temperatureAnalyte Extraction and Purification Protease inhibitor AprotininNo protease inhibitor added
Purpose
The purpose of this study was to evaluate the effects of anticoagulant and up to 6 freeze-thaw cycles at 4 degrees C or RT on cytokine levels. All specimens in this study were spiked with human recombinant cytokines.
Summary of Findings
No obvious differences were observed between specimens thawed at 4 degrees C and those thawed at room temperature. TNF-alpha, IL-1a, IL-1b and IFN-gamma were stable through 6 freeze-thaw cycles regardless of anticoagulant used, however TNF-alpha was significantly lower in serum specimens and IFN gamma was significantly higher in heparinized plasma. IL-6 was stable for at least three freeze-thaw cycles, but was very inconsistent in heparinized plasma.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
FrozenDiagnoses
NormalPlatforms
ProteinRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTAHeparinNoneBiospecimen Aliquots and Components Blood and blood products PlasmaSerumStorage Freeze/thaw cycling 0 cycles1 cycle2 cycles3 cycles4 cycles5 cycles6 cyclesStorage Storage temperature -40 degrees C4 degrees CRoom temperature