Rapid blood separation is superior to fluoride for preventing in vitro reductions in measured blood glucose concentration.
Shi RZ, Seeley ES, Bowen R, Faix JD
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare the use of NaF-KOx and Li-heparin as anticoagulants when determining blood glucose concentrations. Li-heparin specimens underwent centrifugation within 30 minutes, and NaF-KOx specimens were centrifuged 3-8 h post collection. All specimens were refrigerated before analysis.
Summary of Findings
The average blood glucose level was 5.9 mmol/L in NaF-KOx specimens versus 6.1 mmol/L in Li-heparin specimens. Glucose concentrations were an average of 0.39 mmol/L lower in NaF-KOx specimens than in Li-heparin specimens. The authors report no effect of transport time on blood glucose measurements.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Lithium heparinSodium fluoride/potassium oxalateBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Between site transportation method Unspecified transport methodStorage Specimen transport duration/condition 3-8 hAs Li-heparin plasmaNaF-KOx blood (uncentrifuged)