Rapid blood separation is superior to fluoride for preventing in vitro reductions in measured blood glucose concentration.

Shi RZ, Seeley ES, Bowen R, Faix JD

Publication: J Clin Pathol (2009) • Vol 62 • Page 752-3PMID: 19638548View on PubMed

Purpose

The purpose of this paper was to determine if the use of sodium fluoride in blood specimen tubes results in decreased measured glucose concentrations.

Conclusion

Glucose concentrations were an average of 0.39 mmol/L lower in sodium fluoride/potassium oxalate (NaF-KOx) specimens than in Lithium heparin (Li-heparin) specimens.

Studies (1)

Purpose

The purpose of this study was to compare the use of NaF-KOx and Li-heparin as anticoagulants when determining blood glucose concentrations. Li-heparin specimens underwent centrifugation within 30 minutes, and NaF-KOx specimens were centrifuged 3-8 h post collection. All specimens were refrigerated before analysis.

Summary of Findings

The average blood glucose level was 5.9 mmol/L in NaF-KOx specimens versus 6.1 mmol/L in Li-heparin specimens. Glucose concentrations were an average of 0.39 mmol/L lower in NaF-KOx specimens than in Li-heparin specimens. The authors report no effect of transport time on blood glucose measurements.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    Sodium fluoride/potassium oxalate
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageBetween site transportation method
    Unspecified transport method
    StorageSpecimen transport duration/condition
    3-8 h
    As Li-heparin plasma
    NaF-KOx blood (uncentrifuged)