Study of interlaboratory reliability and reproducibility of estrogen and progesterone receptor assays in Europe. Documentation of poor reliability and identification of insufficient microwave antigen retrieval time as a major contributory element of unreliable assays.

Rhodes A, Jasani B, Balaton AJ, Barnes DM, Anderson E, Bobrow LG, Miller KD

Publication: Am J Clin Pathol (2001) • Vol 115 • Page 44-58PMID: 11190807View on PubMed

Purpose

The purpose of this paper was to determine the factors influencing reproducibility of immunohistochemical (IHC) assays for estrogen (ER) and progesterone receptors (PR) in breast cancer tissue.

Conclusion

An increase in average microwave mediated antigen retrieval times from 18 to 26 minutes significantly improved ER and PR IHC staining, regardless of antigen retrieval buffer or pH. Use of a pressure cooker, as opposed to a microwave for antigen retrieval, also appeared to improve IHC staining. Primary antibody conditions such as dilution, incubation time, and temperature did not significantly impact ER and PR IHC staining.

Studies (1)

Purpose

The purpose of this study was to determine the effects of antigen retrieval method and primary antibody incubation conditions on the reproducibility of IHC assays for ER and PR in breast cancer composite tissue blocks.

Summary of Findings

Of reproducible IHC assays, 54% were obtained in laboratories that used pressure cooker-mediated antigen retrieval, and 34% were from laboratories that used a microwave for antigen retrieval. An average of only 26% of laboratories not achieving the reproducibility standards used a pressure cooker, while 60% used microwave assisted antigen retrieval. Reproducible IHC results were obtained by laboratories using a wide variety of antigen retrieval buffers at a range of pH values. Among the laboratories without reproducible IHC staining, increasing microwave antigen retrieval time by 10 minutes, resulting in an average antigen retrieval time of 26 minutes versus 18 minutes, significantly improved IHC reproducibility outcomes, but there was no effect when primary antibody conditions were altered.
  • Biospecimens

    TissueBreast
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticBenign
    NeoplasticCarcinoma
  • Platforms

    ProteinImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationAntigen retrieval
    Various antigen retrieval times
    Various buffers
    Various pH values
    Analyte Extraction and PurificationTemperature of heat-induced retrieval
    Microwave
    Pressure cooker
    Immunohistochemistry SpecificAntibody dilution/concentration
    Various dilutions
    Immunohistochemistry SpecificIncubation temperature
    Various temperatures
    Immunohistochemistry SpecificIncubation time
    Various durations
    Immunohistochemistry SpecificTargeted peptide/protein
    ER
    PR