Storage at -80 degrees C decreases the concentration of HPLC-detected urinary albumin: possible mechanisms and implications.
Markó L, Cseh J, Kószegi T, Szabó Z, Molnár GA, Mohás M, Szigeti N, Wittmann I
Purpose
The purpose of this paper was to determine the effects of storage of urine specimens at -80 degrees C on HPLC-detected urinary albumin concentration (UAC).
Conclusion
Storage of urine at -80 degrees C for 2.5 years resulted in decreased UAC by an average of 24%. Lower urinary pH was associated with a greater decrease in HPLC-detected UAC. The dimeric to monomeric ratio (DMR) of urinary albumin increased in stored specimens with significant changes in the peak area of the monomeric but not the dimeric form. Total free sulfhydryl groups (TFSG) were significantly lower in stored urine specimens when compared to fresh specimens.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage of urine specimens at -80 degrees C on HPLC-detected UAC.
Summary of Findings
Storage of urine at -80 degrees C for 2.5 years resulted in decreased UAC by an average of 24%. Lower urinary pH was associated with a greater decrease in HPLC-detected UAC. The DMR of urinary albumin increased in stored specimens with significant changes in the peak area of the monomeric but not dimeric form. TFSG were significantly lower in stored urine specimens and did not correlate with specimen pH. On the other hand, in fresh specimens, an exponential correlation with urine pH was observed.
Biospecimens
FluidUrinePreservation Types
FrozenDiagnoses
Diabetes Type 2Other diagnosesPlatforms
ProteinHPLCElectrolyte/MetalpHSmall moleculeSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition MicroalbuminuriaNormoalbuminuriaBiospecimen Aliquots and Components pH High pHLow pHStorage Storage duration 2.5 yLess than 2 weeksStorage Storage temperature -80 degrees C