Influence of sampling and storage conditions on plasma renin activity and plasma renin concentration.

Locsei Z, Racz K, Patocs A, Kovacs GL, Toldy E

Publication: Clin Chim Acta (2009) • Vol 402 • Page 203-5PMID: 19263530View on PubMed

Purpose

The purpose of this paper was to determine the optimum sampling and storage conditions when analyzing plasma renin activity (PRA) and concentration (REN).

Conclusion

PRA and REN were highest in specimens collected into prechilled tubes, processed at 0-5 degrees C, and centrifuged within 30 minutes of specimen collection. Storage of plasma at -20 degrees C for 7 weeks caused a significant decrease in PRA (31.9 +/- 4.7%) while REN changed by less than 5%.

Studies (2)

Purpose

The purpose of this study was to determine the optimum blood specimen processing conditions and freezing method for the measurement of PRA and REN.

Summary of Findings

When blood specimens were collected, held, and centrifuged at room temperature after a 2 hour delay in processing, PRA and REN significantly decreased (p<0.001). The same 2 hour delay in processing did not affect the analytes if collection, storage, and centrifugation took place at 0-5 degrees C. Separation of plasma within 30 minutes as opposed to 2 hours resulted in higher REN and PRA. Rapid freezing did not significantly impact REN or PRA levels.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    ProteinRadioimmunoassay
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Prechilled tube
    Room temperature tube
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    Biospecimen PreservationCooling or freezing method/ rate
    Rapid freezing at -80 degrees C
    Standard freezing at -20 degrees C
    StorageStorage duration
    2 h
    30 min
    StorageStorage temperature
    0-5 degrees C
    Room temperature