Effect of sample collection on alpha-galactosidase A enzyme activity measurements in dried blood spots on filter paper.
Olivova P, van der Veen K, Cullen E, Rose M, Zhang XK, Sims KB, Keutzer J, Browning MF
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this paper was to determine the effects of sample collection method on AGAL activity in dried blood spots from normal individuals and patients with Fabry disease. The effects of storage and anticoagulant concentration on AGAL in tube-collected specimens were also examined.
Summary of Findings
No difference in AGAL activity was found with direct spotting of blood collected by venipuncture or lancet. However, AGAL activity was up to 68% higher in specimens collected in EDTA tubes and spotted 24-72 h later (p=0.0007). Storage duration and EDTA concentration did not significantly effect AGAL activity in tube-collected blood specimens, but the coefficient of variance (CV) was higher in specimens stored for 72 h prior to spotting than in those stored for 24 h. In Fabry disease patients, AGAL activity was significantly lower. No significant effects of gender or age were observed, but some females had levels that overlapped the normal range. In conclusion, AGAL activity is dependent on collection method, and in blood specimens collected in tubes, anticoagulant concentration and storage for up to 72 h do not influence activity.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesPlatforms
ProteinEnzyme assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Fabry diseaseNormalPreaquisition Patient gender FemaleMaleBiospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Method of fluid acquisition Finger/heel prick samplingVenipunctureBiospecimen Acquisition Type of collection container/solution EDTA vacutainer tubeFilter paperBiospecimen Aliquots and Components Aliquot size/volume 10 mL2 mL5 mLStorage Storage duration 24 h48 h72 hStorage Storage temperature 4 degrees C