Variation in plasma potassium when using collection tubes containing gel separators.
Jowitt S, Longlands MG, Probert DE
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine whether the use of gel separator tubes and increased time at room temperature before separation of blood specimens leads to alterations in measured plasma potassium levels.
Summary of Findings
Collection of blood into gel separator tubes did not cause falsely elevated plasma potassium levels unless separation was delayed, and patient had a high white blood cell count. It is important to note that this combination of variables was observed for only one leukemia patient. Nine other hematology clinic patients with white blood cell counts in the normal range did not show falsely elevated plasma potassium levels as a result of delayed separation and/or the use of gel separator tubes.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Other diagnosesNeoplasticLeukemiaPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level High white blood cell countNormal white blood cell countBiospecimen Acquisition Anticoagulant HeparinBiospecimen Acquisition Type of collection container/solution Tube containing a gel separatorTube with no gel separatorBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 1 h10 minSeveral hours