Evaluation of effect of specimen-handling parameters for plasma preparation tubes on viral load measurements obtained by using the Abbott RealTime HIV-1 load assay.

Fernandes H, Morosyuk S, Abravaya K, Ramanathan M, Rainen L

Publication: J Clin Microbiol (2010) • Vol 48 • Page 2464-8PMID: 20484602View on PubMed

Purpose

The purpose of this paper was to determine the effects of delayed centrifugation and freezing plasma in situ after centrifugation on Human immunodeficiency virus (HIV) load determination.

Conclusion

No effect of delayed centrifugation was found on HIV load determination. K2 EDTA tubes and plasma preparation tubes performed similarly. Transfer of plasma specimens to new tubes prior to or after freezing did not effect viral load determination. Variability between measurements was found to be increased in specimens with a viral load below 100 copies/ml, but no effect of specimen handling was found. In conclusion, specimens collected in plasma preparation tubes can be stored for 6 h prior to centrifugation and can be frozen in the same tube without affecting viral load determination.

Studies (1)

Purpose

The purpose of this study was to determine the effects of delayed centrifugation and freezing plasma in situ after centrifugation on HIV load determination

Summary of Findings

No effect of delayed centrifugation was found on HIV load determination. K2 EDTA tubes and plasma preparation tubes performed similarly as did specimens which were aliquoted prior to and after freezing. Variability between measurements was found to be increased in specimens with a viral load below 100 copies/ml, but no effect of specimen handling was found. In conclusion, specimens collected in plasma preparation tubes can be stored for 6 h prior to centrifugation and can be frozen in the same tube without affecting viral load determination.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    AIDS/HIV-related
  • Platforms

    RNAReal-time qRT-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPrognostic factor
    50-5000 viral copies/ml
    Less than 50 viral copies/ml
    Biospecimen AcquisitionType of collection container/solution
    BD Vacutainer K2 EDTA tube
    BD Vacutainer plasma preparation tube
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage temperature
    -70 degrees C
    StorageTime at room temperature
    2 h
    6 h
    StorageType of storage container
    New tube
    Original collection tube