Evaluation of effect of specimen-handling parameters for plasma preparation tubes on viral load measurements obtained by using the Abbott RealTime HIV-1 load assay.
Fernandes H, Morosyuk S, Abravaya K, Ramanathan M, Rainen L
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of delayed centrifugation and freezing plasma in situ after centrifugation on HIV load determination
Summary of Findings
No effect of delayed centrifugation was found on HIV load determination. K2 EDTA tubes and plasma preparation tubes performed similarly as did specimens which were aliquoted prior to and after freezing. Variability between measurements was found to be increased in specimens with a viral load below 100 copies/ml, but no effect of specimen handling was found. In conclusion, specimens collected in plasma preparation tubes can be stored for 6 h prior to centrifugation and can be frozen in the same tube without affecting viral load determination.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
FrozenDiagnoses
AIDS/HIV-relatedPlatforms
RNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Prognostic factor 50-5000 viral copies/mlLess than 50 viral copies/mlBiospecimen Acquisition Type of collection container/solution BD Vacutainer K2 EDTA tubeBD Vacutainer plasma preparation tubeBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage temperature -70 degrees CStorage Time at room temperature 2 h6 hStorage Type of storage container New tubeOriginal collection tube