Stability of lyophilised specimens for the molecular detection of viral DNA/RNA.
Vaughan H, Chalker VJ, Mee Z, Rossouw A, James V
Purpose
Conclusion
Studies (3)
Purpose
The purpose of this study was to determine whether hepatitis B viral load is affected by storage of infected lyophilized serum at room temperature, 4 degrees C, or -30 degrees C.
Summary of Findings
No decrease in viral load was detected using either a manual PCR assay or an automated one after 11 months of storage at room temperature, 4 degrees C, or -30 degrees C.
Biospecimens
FluidSerumPreservation Types
Other PreservativeDiagnoses
HepatitisPlatforms
DNAPCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Hepatitis BStorage Storage duration 1 month10 d11 months2 months3 months4 months5 months6 monthsStorage Storage temperature -30 degrees C4 degrees CRoom temperaturePCR Specific Technology platform Amplicor COBAS Monitor assay (automated)Roche Amplicor Monitor assay (manual)
Purpose
The purpose of this study was to determine whether hepatitis C viral load is affected by storage of infected lyophilized plasma at room temperature or -30 degrees C.
Summary of Findings
Storage of lyophilized specimens at room temperature decreased the hepatitis C viral load by 0.74 log10 copies/mL in one specimen after 7 months and by 0.24 log10 copies/mL in the second specimen after 6 months. Storage of lyophilized plasma specimens for up to 26 months at -30 degrees C had no effect on hepatitis C viral load.
Biospecimens
FluidPlasmaPreservation Types
Other PreservativeDiagnoses
HepatitisPlatforms
RNART-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Hepatitis CStorage Storage duration 10 d15 months2 months26 months6 months7 monthsStorage Storage temperature -30 degrees CRoom temperature
Purpose
The purpose of this study was to determine whether HIV viral load is affected by storage of lyophilized infected plasma at -30 degrees C.
Summary of Findings
After lyophilized plasma specimens were stored for 36 months at -30 degrees C there was no significant decrease detected in HIV viral load by the Bayer branch DNA method. However, when the Roche COBAS Amplicor Monitor PCR-based assay was used, decreases in viral load up to 0.51 log10 copies/mL were detected, but the authors suggest this was due to inter-assay variability rather than a significant loss in viral load.
Biospecimens
FluidPlasmaPreservation Types
Other PreservativeDiagnoses
AIDS/HIV-relatedPlatforms
RNART-PCRRNABranch DNA assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Prognostic factor Viral load 10,000 copies/mLViral load 1000 copies/mLViral load 40,000 copies/mLViral load 4000 copies/mLStorage Storage duration 10 d36 monthsStorage Storage temperature -30 degrees C