Artificial prolongation of the prothrombin time of lyophilized plasma induced by transportation with solid carbon dioxide.
Vis R, van den Besselaar AM, Witteveen E, Meeuwisse-Braun J, Schaefer-van Mansfeld H
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effect of shipping of lyophilized plasma with dry ice on prothrombin time and specimen pH.
Summary of Findings
Lyophilization of plasma specimens increased reconstituted specimen pH from 7.1 to 7.7. Shipping of the lyophilized plasma on dry ice versus without dry ice increased the average prothrombin time from 112.7 s to 117.8 s and decreased specimen pH. In addition the inter-vial variability increased from 1% in specimens shipped without dry ice to 5% in specimens shipped on dry ice. The specimens with the longest prothrombin time had the lowest pH indicating the increased prothrombin time and variability may be due to leakage of carbon dioxide into some specimens. The authors recommend shipping of lyophilized plasma specimens for coagulation assays without dry ice or with special care to avoid carbon dioxide contamination.
Biospecimens
FluidPlasmaPreservation Types
Other PreservativeDiagnoses
NormalOther diagnosesPlatforms
Cell count/volumeHematology/ auto analyzerSmall moleculepHElectrolyte/MetalColorimetric assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition HaemophiliaPreaquisition Other drugs Oral anticoagulantsStorage Storage conditions With dry iceWithout dry ice