Comparison of glass K3EDTA versus plastic K2EDTA blood-drawing tubes for complete blood counts, reticulocyte counts, and white blood cell differentials.
Van Cott EM, Lewandrowski KB, Patel S, Grzybek DY, Patel HS, Fletcher SR, Kratz A
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare collection and storage of blood specimens in glass K3EDTA tubes versus plastic K2EDTA tubes for complete blood and reticulocyte counts, white blood cell differentials, and flagging rates on the automated cell counter.
Summary of Findings
Minor statistically significant differences (less than 5%) in some parameters were seen between the two collection tube types both before and after overnight storage at 4 degrees C. Specimens in Vacutainer K3EDTA glass tubes had slightly lower values than those in Vacutainer K2EDTA Plus plastic tubes. However, the authors believe none were likely to be of clinical relevance. The authors attribute some of the discrepancies to the liquid K3EDTA anticoagulant which causes specimen dilution versus the spray-coating of dry K2EDTA. There was no difference in flagging rates on the automated cell counter.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesNeoplasticOtherPlatforms
Cell count/volumeHematology/ auto analyzerProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Multiple forms evaluatedPotassium EDTABiospecimen Acquisition Type of collection container/solution Glass tubePlastic tubeStorage Storage duration 24 h4 hStorage Storage temperature 4 degrees C