Problems with p53 immunohistochemical staining: the effect of fixation and variation in the methods of evaluation.
Fisher CJ, Gillett CE, Vojtesek B, Barnes DM, Millis RR
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the optimal fixative choice and fixation temperature for p53 immunohistochemistry (IHC) staining of breast carcinomas.
Summary of Findings
Specimens fixed with phenol formol saline at room temperature had the clearest morphology and p53 staining. The next best staining was seen with formol saline fixation at 4 degrees or methacarn fixation at room temperature. The least satisfactory p53 staining was observed in specimens preserved in formol calcium at any temperature or formol saline used at 55 degrees C. Background staining was a problem in specimens fixed with neutral buffered formalin or neutral buffered phenol formalin. IHC of specimens fixed with phenol formol saline showed a high degree of correlation with ELISA scores (r=0.871, p<0.0001). Increased p53 staining intensity was observed with increasing cancer grade. In conclusion, phenol formol saline is the best fixative choice for analysis of breast carcinomas by H&E staining and p53 IHC.
Biospecimens
TissueBreastPreservation Types
FormalinOther PreservativeDiagnoses
NeoplasticCarcinomaPlatforms
MorphologyH-and-E microscopyProteinELISAProteinImmunohistochemistryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Temperature of fixation/preservation 4 degrees C55 degrees CRoom temperatureBiospecimen Preservation Type of fixation/preservation Formalin (buffered)Formol calciumFormol salineMethacarn