Measurements of free hemoglobin and hemolysis index: EDTA- or lithium-heparinate plasma?
Unger J, Filippi G, Patsch W
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine whether EDTA or lithium heparinate is the preferable anticoagulant for measuring fHb and HI in plasma.
Summary of Findings
No significant differences in fHb or HI were observed between the two anticoagulants when they were used at usual concentrations (1.6 g EDTA/L of whole blood or 16 kIE lithium heparinate/L of whole blood). However, when EDTA was used at 3 times the normal concentration, as occurs when collection tubes are not adequately filled, fHb and HI were over 50% higher than when EDTA was used at standard concentrations. On the other hand, increasing lithium heparinate concentration did not affect fBh or HI.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalNot specifiedPlatforms
ProteinSpectrophotometryCell count/volumeHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTALithium heparin