Skip to main content

Insufficient filling of vacuum tubes as a cause of microhemolysis and elevated serum lactate dehydrogenase levels. Use of a data-mining technique in evaluation of questionable laboratory test results.

Tamechika Y, Iwatani Y, Tohyama K, Ichihara K

Publication: Clin Chem Lab Med (2006) • Vol 44 • Page 657-61PMID: 16681441View on PubMed

Purpose

The purpose of this paper was to determine the effects of partially filling vacuum specimen tubes as well as time prior to centrifugation on lactate dehydrogenase (LDH) levels, hemolysis, and other routine biochemical screening test analytes.

Conclusion

Exposure of blood specimens to unfilled vacuum in collection tubes prior to centrifugation significantly increased LDH, aspartate aminotransferase (AST), and potassium (K) levels as well as hemolysis. LDH levels and hemolysis were found to increase directly with duration of exposure. Exposure of specimens to unfilled vacuum during centrifugation did not increase hemolysis.

Studies (2)

Purpose

The purpose of this study was to determine the effects of unfilled vacuum in blood specimen collection tubes on hemolysis, LDH, alanine aminotransferase (ALT), AST, gamma-glutamyl transpeptidase (GGT), sodium (Na), K, and chloride.

Summary of Findings

Levels of LDH, AST, and K increased by 8.0%, 3.8%, and 3.4%, respectively in serum when sample tubes had unfilled vacuum compared to those with no unfilled vacuum. Unfilled vacuum in collection tubes had no significant effect on ALT, GGT, Na, or chloride. When hemolysis was semi-quantitatively measured by urine stick analysis, 8/10 paired specimens showed a significantly higher hemoglobin concentration in the sample tube containing unfilled vacuum.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    MorphologyColorimetric assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    2 mL (with a 9 mL draw volume)
    6 mL (with a 6 mL draw volume)
    StorageTime at room temperature
    90 min