Skip to main content

Effect of collection tube type and preanalytical handling on myeloperoxidase concentrations.

Shih J, Datwyler SA, Hsu SC, Matias MS, Pacenti DP, Lueders C, Mueller C, Danne O, Möckel M

Publication: Clin Chem (2008) • Vol 54 • Page 1076-9PMID: 18509013View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type, storage duration prior to and following centrifugation, storage temperature, and freeze-thaw cycling on mycloperoxidase (MPO) concentration.

Conclusion

EDTA or EDTA plasma preparation tubes consistently performed the best with respect to stabilizing MPO concentration. Whole blood was able to be stored at room temperature for 2 hours or on ice for 8 hours in these tubes without significantly affecting MPO concentrations. In addition, separated plasma was able to be stored at room temperature for 2 days, at 2-8 degrees C for 8 days, or through 3 freeze/thaw cycles in these tubes without significantly affecting MPO concentration.

Studies (3)

Purpose

The purpose of this study was to determine the effects of collection tube type and storage duration prior to centrifugation either in an ice bath or at room temperature on MPO concentration in normal blood specimens.

Summary of Findings

EDTA and citrate plasma tubes gave the lowest concentrations of MPO and were not significantly different. Average MPO concentrations were significantly higher in heparin (10%) and serum specimens (up to 100%) (p<0.001). A spike recovery experiment showed that the higher MPO concentrations were not due to direct anticoagulant interference. MPO concentration was stable when whole blood was kept on ice for 8 hours or at room temperature for 2 hours prior to centrifugation in either EDTA or EDTA plasma preparation tubes. In citrate and lithium heparin tubes, MPO concentrations were only stable when whole blood was kept on ice for 2 hours prior to centrifugation and were not stable when whole blood was stored at room temperature.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    K2 EDTA
    K2 EDTA plasma preparation
    Lithium heparin (glass)
    Lithium heparin (plastic)
    Lithium heparin plasma separator
    Serum (glass)
    Serum (plastic)
    Serum separator
    Sodium citrate
    Sodium heparin
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 h
    2 h
    8 h
    StorageStorage temperature
    0 degrees C
    RT