Effect of time, temperature and additives on a functional assay of C1 inhibitor.

Nielsen EW, Johansen HT, Straume B, Mollnes TE

Publication: J Immunol Methods (1994) • Vol 173 • Page 245-51PMID: 8046257View on PubMed

Purpose

The purpose of this paper was to determine the optimal conditions for blood specimen collection and storage when diagnosing hereditary angioedema (HAE) through measurement of the C1 inhibitor (C1-INH).

Conclusion

C1-INH activity was elevated in heparin treated plasma and depressed in polybrene-EDTA treated plasma. Significant declines in C1-INH levels were found in specimens stored at 4 degrees C and 37 degrees C. The decline in C1-INH activity with storage at either temperature was more pronounced in HAE patients than in normal patients. Storage of serum for 5 days at room temperature reduced C1-INH activity, but an observable effect on C1-INH levels by Western blot was only found in HAE patients. EDTA-treated plasma or serum were the best choices for measurement of C1-INH activity and while C1-INH activity levels declined with storage, HAE and normal specimens were still distinguishable after storage of serum for 5 days at room temperature.

Studies (2)

Purpose

The purpose of this study was to determine the effects of anticoagulant and storage at 4 or 37 degrees C for up to 24 h on the activity of C1-INH in plasma and serum from normal and HAE patients. Serum specimens were allowed to coagulate for 2 h at room temperature prior to centrifugation. After the allotted storage time, all specimens were stored at -70 degrees C until analysis.

Summary of Findings

C1-INH activity was falsely elevated in heparin treated plasma and falsely depressed in polybrene-EDTA treated plasma. Significantly higher C1-INH activity was observed in serum, EDTA-treated plasma and heparin-treated plasma, then in citrate-treated plasma from healthy volunteers (p=0.0003-0.0025), but not from HAE patients (p=0.055-0.8). Activity of C1-INH decreased by 12-19% after storage of normal plasma (EDTA, heparin, or citrate-treated) or serum for 24 h at 37 degrees C. A decline in C1-INH activity of 5-14% was noted after 24 h at 4 degrees C for normal plasma (EDTA, heparin, or citrate-treated) or serum. The decline in C1-INH activity was far more pronounced in HAE patients after storage at either 37 (48-83%) or 4 degrees C (35-80%) and was also found for polybrene-EDTA-treated plasma. Although a significant effect of storage was found for both healthy volunteers (p=0.039) and HAE patients (0.0002) there was no significant difference between the two temperatures. In conclusion, EDTA-treated plasma or serum were the best choices for measurement of C1-INH activity and C1-INH activity declines with storage at 4 or 37 degrees C.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    ProteinEnzyme assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Hereditary angioedema
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Heparin
    Polybrene EDTA
    Sodium citrate
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    StorageStorage conditions
    0 h
    24 h
    6 h
    StorageStorage temperature
    -70 degrees C
    37 degrees C
    4 degrees C