Effects of prolonged ambient storage of sodium fluoride/heparin specimens on plasma homocysteine.
Scheidhauer R, Guessregen B, Hohl A, Arndt T
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare homocysteine levels in blood collected in tubes with the anticoagulant NaF/heparin and stored for up to 48 h at room temperature prior to centrifugation with specimens collected in tubes containing EDTA and centrifuged after 15 minutes.
Summary of Findings
Specimens collected in tubes containing EDTA and centrifuged within 15 minutes and specimens collected in tubes with NaF/heparin and stored for 24-144 h prior to centrifugation had equivalent levels of homocysteine. Specimens collected in NaF/heparin tubes and centrifuged within 15 minutes had significantly lower levels of homocysteine than EDTA specimens centrifuged within 15 minutes. This was presumably due to plasma dilution from erythrocyte water loss caused by the NaF, which was no longer an issue at the 24-hour time point. In conclusion, the use of NaF/heparin allows for storage of blood for up to 144 h at room temperature prior to centrifugation without altered homocysteine levels.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Small moleculeLC-MS or LC-MS/MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTASodium fluoride/heparinBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Time at room temperature 144 h15 min24 h48 h