Avoidance of contamination in element analysis of serum samples.
Lakomaa EL, Mussalo-Rauhamaa H, Salmela S
Purpose
The purpose of this paper was to determine the effects of long term storage of serum at -18 degrees C on copper, zinc, selenium, sodium, and chloride concentrations after repeated freeze-thaw cycles.
Conclusion
There were no statistically significant changes in copper zinc, or selenium concentrations after storing serum for over 4 years at -18 degrees C. There was also no change in sodium or chloride concentration over 29 months, which the authors believe indicates no water loss.
Studies (1)
Purpose
The purpose of this study was to determine the effects of long term storage of serum at -18 degrees C on copper, zinc, selenium, sodium, and chloride concentrations after repeated freeze-thaw cycles. Polyethylene storage tubes were prewashed with 1% Deconex overnight, and then a 1+1 solution of suprapure HNO3.
Summary of Findings
There were no statistically significant changes in copper zinc, or selenium concentrations after storing serum for over 4 years at -18 degrees C. There was also no change in sodium or chloride concentration over 29 months, which the authors believe indicates no water loss.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
Electrolyte/MetalAtomic absorption spectroscopyElectrolyte/MetalInstrumental neutron activation analysisPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Freeze/thaw cycling 0 cycles1 cycle2 cycles3 cycles4 cyclesStorage Storage duration 0 months19 months29 months34 months43 months46 months51 monthsStorage Storage temperature -18 degrees C