Spuriously positive direct antiglobulin tests caused by use of silicone gel.

Geisland JR, Milam JD

Publication: Transfusion (1980) • Vol 20 • Page 711-3PMID: 7434455View on PubMed

Purpose

The purpose of this paper was to assess the suitability of silicone gel containing tubes for anti-globulin tests in blood bank and transfusion services.

Conclusion

Red blood cell agglutination was observed macroscopically in 13% of specimens collected in serum separator tubes, but not at all in nonsiliconized tubes. Although agglutination occurred in both Corvac and Vaccutainer tubes, there were 5 additional positives from specimens collected in Corvac tubes. The agglutination in the positive specimens was caused by anti-C3, but not IgG. In conclusion, silicone gel containing tubes should be used for clinical chemistry but not for blood bank and transfusion service analysis.

Studies (1)

Purpose

The purpose of this study was to determine the effects of silicone gel containing tubes on aggregation of red blood cells in anti-globulin (C3 and IgG) tests. Specimens were refrigerated.

Summary of Findings

Red blood cell agglutination was observed macroscopically in 13% of specimens collected in serum separator tubes but not at all in nonsiliconized tubes. Although agglutination occurred in both Corvac and Vaccutainer tubes, there were 5 additional positives from specimens collected in Corvac tubes. The agglutination in the positive specimens was caused by anti-C3, but not IgG. In conclusion silicone gel containing tubes should be used for clinical chemistry but not for blood bank and transfusion service analysis.

  • Biospecimens

    FluidBlood
  • Preservation Types

    Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    MorphologyMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Corvac integrated serum separator tube
    Vacutainer plain tube
    Vacutainer serum separator tube
    Macroscopic observation SpecificTargeted peptide/protein
    Anti-C3
    Anti-IgG