Spuriously positive direct antiglobulin tests caused by use of silicone gel.
Geisland JR, Milam JD
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of silicone gel containing tubes on aggregation of red blood cells in anti-globulin (C3 and IgG) tests. Specimens were refrigerated.
Summary of Findings
Red blood cell agglutination was observed macroscopically in 13% of specimens collected in serum separator tubes but not at all in nonsiliconized tubes. Although agglutination occurred in both Corvac and Vaccutainer tubes, there were 5 additional positives from specimens collected in Corvac tubes. The agglutination in the positive specimens was caused by anti-C3, but not IgG. In conclusion silicone gel containing tubes should be used for clinical chemistry but not for blood bank and transfusion service analysis.
Biospecimens
FluidBloodPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
MorphologyMacroscopic observationPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Corvac integrated serum separator tubeVacutainer plain tubeVacutainer serum separator tubeMacroscopic observation Specific Targeted peptide/protein Anti-C3Anti-IgG