Evaluation of different RNA extraction methods and storage conditions of dried plasma or blood spots for human immunodeficiency virus type 1 RNA quantification and PCR amplification for drug resistance testing.

Monleau M, Montavon C, Laurent C, Segondy M, Montes B, Delaporte E, Boillot F, Peeters M

Publication: J Clin Microbiol (2009) • Vol 47 • Page 1107-18PMID: 19193835View on PubMed

Purpose

The purpose of this paper was to evaluate the effects of viral RNA extraction method and different storage conditions on human immunodeficiency virus (HIV) load determination from dried blood spots (DBS) and dried plasma spots (DPS).

Conclusion

Although all kits performed similarly with fresh plasma, only the Abbot sample preparation system (Abbot) and the Nuclisens manual extraction kit (Nuclisens) resulted in efficient extraction from spiked DPS and patient DPS and DBS. However, the Abbot kit resulted in a slight underestimation of viral load when DBS were analyzed. Storage of DPS for up to 2 months at 20 degrees C did not result in any change in viral load. Storage of DPS at 37 degrees C for 1 month resulted in decreased viral load. Importantly, after only 1 week at 20 degrees C or 37 degrees C, PCR amplification of long fragments was impaired.

Studies (3)

Purpose

The purpose of this study was to evaluate different extraction kits for isolation of HIV from fresh plasma and DPS. Both spiked plasma and patient specimens were used for extraction with the QIAamp viral RNA mini kit (QIAamp), Abbot, Nuclisens, and High Pure viral nucleic acid kit (HighPure).

Summary of Findings

Using fresh plasma specimens, variability between extraction methods was similar to interassay variability. Abbot and Nucliesens kits performed well for all DPS viral loads tested, while HighPure and QIAamp performed poorly with low viral load specimens. All kits had some undetected virus in DPS, but this was particularly a problem for the QIAamp and HighPure kits. PCR amplification of the PR and RT regions of the polymerase gene worked well for the Nuclisens and Abbot kits, but had a high level of failure with the QIAamp and HighPure kits. Results using DPS from patients were similar to those from spiked plasma. In conclusion while no difference between extraction kits were identified when fresh plasma was used, QIAamp and HighPure extraction kits did not perform as well as Abbot and Nuclisens kits when using DPS.
  • Biospecimens

    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    AIDS/HIV-related
  • Platforms

    RNART-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPrognostic factor
    3.3 log(10) viral copies/mL
    4.3 log(10) viral copies/mL
    5.3 log(10) viral copies/mL
    Biospecimen PreservationDehydration duration/condition
    Dried plasma spot
    Fresh plasma
    Analyte Extraction and PurificationAnalyte isolation method
    Abbot sample preparation system (Abbot)
    HighPure viral nucleic acid kit (Roche)
    Nuclisens manual extraction kit (BioMerieux)
    QIAamp viral RNA mini kit (Qiagen)