Stability of common analytes in urine refrigerated for 24 h before automated analysis by test strips.

Froom P, Bieganiec B, Ehrenrich Z, Barak M

Publication: Clin Chem (2000) • Vol 46 • Page 1384-6PMID: 10973869View on PubMed

Purpose

The purpose of this paper was to determine the effect of refrigeration for 24 h on the stability of common urine analytes and to examine the precision of autoanalyzer measurements.

Conclusion

Refrigeration of the specimen for 24 h did not impact the stability of nitrates, ketones, or glucose. Refrigeration did lead to some false test results for red blood cells, leukocytes esterase and proteinuria. While generally the tests were highly precise, at lower levels glucose and proteinuria tests led to false results. In conclusion, automated urinalysis is generally precise, but results may be altered if specimens are refrigerated for 24 h.

Studies (1)

Purpose

The purpose of this study was to determine the effect of refrigeration for 24 h on the stability of 6 common urine analytes measured by autoanalyzer. The precision of the tests was also examined through repeat testing within 1 h.

Summary of Findings

Refrigeration of the specimen for 24 h did not impact the stability of nitrates, ketones, or glucose. Refrigeration led to false-negative test results for red blood cells (p<0.05), and a 25% false negative result rate for leukocytes esterase (p<0.005). Proteinuria at low concentrations (less than 5,000 mg/L) showed an increase in false positive test results (P<0.05) after refrigeration. While the tests for nitrates, ketones, red blood cells and leukocyte esterase showed a high degree of precision, glucose and proteinuria tests led to false results at lower concentrations (less than 10,000 mg/L and less than 5,000 mg/L, respectively). The glucose and proteinuria results may have been affected by a small sample size in the high concentration categories. In conclusion, automated urinalysis is generally precise but results may be altered if specimens are stored for 24 h.

  • Biospecimens

    FluidUrine
  • Preservation Types

    None (Fresh)Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    None (fresh)
    Refrigeration
    StorageStorage duration
    0 h
    24 h