Stability of common analytes in urine refrigerated for 24 h before automated analysis by test strips.
Froom P, Bieganiec B, Ehrenrich Z, Barak M
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effect of refrigeration for 24 h on the stability of 6 common urine analytes measured by autoanalyzer. The precision of the tests was also examined through repeat testing within 1 h.
Summary of Findings
Refrigeration of the specimen for 24 h did not impact the stability of nitrates, ketones, or glucose. Refrigeration led to false-negative test results for red blood cells (p<0.05), and a 25% false negative result rate for leukocytes esterase (p<0.005). Proteinuria at low concentrations (less than 5,000 mg/L) showed an increase in false positive test results (P<0.05) after refrigeration. While the tests for nitrates, ketones, red blood cells and leukocyte esterase showed a high degree of precision, glucose and proteinuria tests led to false results at lower concentrations (less than 10,000 mg/L and less than 5,000 mg/L, respectively). The glucose and proteinuria results may have been affected by a small sample size in the high concentration categories. In conclusion, automated urinalysis is generally precise but results may be altered if specimens are stored for 24 h.
Biospecimens
FluidUrinePreservation Types
None (Fresh)Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation None (fresh)RefrigerationStorage Storage duration 0 h24 h