Clinical validation of a new blood collection tube for the accuracy of total homocysteine measurement by different methods.

Bissé E, Epting T, Kögel G, Obeid R, Gempel K, Huzly D, Fuhrmann K, Rolinski B, Herrmann W, Wieland H

Publication: Clin Biochem (2007) • Vol 40 • Page 739-43PMID: 17467680View on PubMed

Purpose

The purpose of this paper was to evaluate the effectiveness of Homocysteine-Primavette tubes at stabilizing total homocysteine (tHcy) levels in plasma as measured by fluorescence polarization immunoassay (FPIA), GC-MS, LC-MS, chemiluminescence immunoassay (ICL), and HPLC.

Conclusion

Collection of blood into Homocysteine-Primavette tubes prevented falsely elevated tHcy concentrations for the duration of the experiments (40 h) while tubes containing EDTA alone or EDTA plus sodium fluoride as a stabilizer did not. However, use of a short cyano column for LC-MS resulted in values above the baseline tHcy value, taken as a reference value, in 60% of the stabilized samples indicating this specific method of tHcy measurement may be incompatible with the use of Homocysteine-Primavette tubes.

Studies (1)

Purpose

The purpose of this study was to compare tHcy concentrations, using tubes containing three different types of collection media, at various time points after blood specimen collection. Five methods of tHcy measurement were employed and average differences between the baseline tHcy and tHcy at 40 h was calculated for stabilized specimens.

Summary of Findings

Baseline tHcy concentration was measured after immediate centrifugation of EDTA only tubes (Primavette) and used as a reference value. While Homocysteine-Primavette tubes were able to stabilize tHcy concentration over the course of 40 h, tHcy concentration increased dramatically (5 fold) in Primavette tubes. A continuous increase in tHcy concentration was seen in Monovette tubes (sodium fluoride + EDTA) up to 6 h at which point the concentration declined. Importantly, baseline tHcy concentration in Monovette tubes was 17% lower than in the other two tube types. Homocysteine-Primavette tubes proved compatible with FPIA, ICL, GCMS and HPLC methods as the average differences between reference value tHcy (0 h) and tHcy concentration at 40 h resulted in mean proportional biases ranging from only -2.7% to 2.0% depending on the method. However, LCMS showed a mean proportional bias of -14% suggesting the stabilizer in Homocysteine-Primavette tubes is incompatible with this method.

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Cardiovascular Disease
    Normal
  • Platforms

    Small moleculeGC-MS
    Small moleculeHPLC
    Small moleculeImmunoassay
    Small moleculeLC-MS or LC-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Sodium fluoride/EDTA
    Biospecimen AcquisitionType of collection container/solution
    Homocysteine-Primavette tube (stabilizer + EDTA)
    Monovette tube (sodium fluoride + EDTA)
    Primavette tube (EDTA)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage temperature
    -20 degrees C
    StorageTime at room temperature
    0 h
    24 h
    40 h
    5 h