Plastic versus siliconized glass tubes: evaluation in current laboratory practice.
Biron-Andréani C, Mallol C, Séguret F, Schved JF
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare the results of several coagulation analyses between specimens collected in two different types of tubes. Analyses performed included activated partial thromboplastin time test (APTT), PT, fibrinogen, anti-Xa activity, factor XIIa (FXIIa) test, and prothrombin fragments 1 + 2 testing.
Summary of Findings
No significant differences were seen with anti-Xa testing or APTT between the two types of tubes with the exception of APTT in patients receiving oral anti-coagulants. In this case, APTT was longer (49.2 s vs 48.5 s) in the siliconized glass tubes (p=0.01). Significant differences between the two collection tube types were also seen in PT and prothrombin fragments 1 + 2 testing for healthy individuals and those receiving oral anti-coagulants or heparin treatment, although the magnitude and direction of tube-mediated effect was dependent on patient diagnosis and the type of anti-coagulant administered. While average fibrinogen and FXIIa concentrations were significantly lower using the siliconized glass tubes for healthy individuals (p<0.005 and p=0.01), there was no significant difference for individuals taking oral anticoagulants.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)FrozenDiagnoses
NormalOther diagnosesPlatforms
GlycoproteinHematology/ auto analyzerProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition HealthyLow molecular weight heparin therapyOral anticoagulant therapyUnfractionated heparin therapyBiospecimen Acquisition Anticoagulant CitrateBiospecimen Acquisition Type of collection container/solution Plastic tubeSiliconized glass tubeBiospecimen Preservation Type of fixation/preservation None (fresh)Snap frozenStorage Storage temperature -80 degrees CRT