Stability of free prostate-specific antigen in serum samples under a variety of sample collection and sample storage conditions.

Woodrum D, French C, Shamel LB

Publication: Urology (1996) • Vol 48 • Page 33-9PMID: 8973697View on PubMed

Purpose

The purpose of this paper was to determine the stability of both free and total prostate specific antigen (PSA) in blood specimens after different collection and storage conditions were employed.

Conclusion

While total PSA was generally stable throughout all the studies, free PSA decreased by approximately 1% per hour of clotting time, by 2-3% per day of storage at 4 degrees C or 23 degrees C, and by 0.9% and 0.4% per month of storage at -20 degrees C and -70 degrees C respectively. Freeze-thaw cycles did not impact total or free PSA values. Consequently, the authors conclude that routine collection, preparation, and refrigeration of samples for up to 24 hours should not significantly impact the measurements of free and total PSA. Freezing samples for longer storage is advised.

Studies (4)

Purpose

The purpose of this study was to characterize the impact of allowing blood to clot at RT for different amounts of time before preparing serum samples for total and free PSA value determination. For this study, blood specimens were obtained from 11 different subjects, and serum was snap-frozen at -70 degrees C.

Summary of Findings

While total PSA values stayed the same, free PSA values decreased at an average rate of 1% per hour of clotting time at RT (P=0.0001).

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    Frozen
  • Diagnoses

    Prostatitis
    NeoplasticCarcinoma
    NeoplasticBenign
  • Platforms

    GlycoproteinImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageTime at room temperature
    1 h
    3 h
    5 h
    8 h