Estimation of the minimal preanalytical uncertainty for 15 clinical chemistry serum analytes.

Sylte MS, Wentzel-Larsen T, Bolann BJ

Publication: Clin Chem (2010) • Vol 56 • Page 1329-35PMID: 20551383View on PubMed

Purpose

The purpose of this paper was to identify the contribution of collection tube type, clotting time, arm used for collection, order of sequential collection and measurement repeatability on the variability of 15 serum analytes

Conclusion

No significant effect was observed with regard to the arm used for collection or the order of sequential collection. However mean values obtained for albumin, calcium, glucose, lactate dehydrogenase (LDH), magnesium, potassium and total cholesterol differed significantly between serum-separation tubes (SST)-II advance and rapid serum tubes (4.0) (RST). Hemolysis was also significantly higher in the SST tubes. The authors attribute these differences to different clotting times and coagulant agents associated with tube type. Between venipunctures significant variation was found in glucose, alkaline phosphatase and potassium. The authors note that small physiological changes such as arm position, blood flow, and blood pressure as well as changes between venipunctures in depth, tourniquet strength, and duration may all be responsible. While measurement repeatability was the lowest achievable, it was the largest contributor to standard deviation for gamma-glutamyltransferase, alanine aminotransferase and magnesium. The authors conclude that there are many confounding factors to consider when analyzing serum analytes including, those of preprocessing (tube type, clotting time, coagulant), collection (arm position, tourniquet strength, venipuncture depth and duration) and processing (extent of hemolysis, mixing time, time before analysis, and room temperature and humidity) as well as error introduced in measurement.

Studies (1)

Purpose

The purpose of this study was to identify the contribution of collection tube type, clotting time, arm used for collection, order of sequential collection and measurement repeatability on the variability of 15 serum analytes.

Summary of Findings

No significant effect was observed with regard to the arm used for collection or the order of sequential collection. However mean values obtained for albumin, calcium, glucose, lactate dehydrogenase (LDH), magnesium, potassium and total cholesterol differed significantly between serum-separation tubes (SST)-II advance and rapid serum tubes (4.0)(RST). Hemolysis was also significantly higher in the SST tubes. The authors attribute these differences to different clotting times and coagulant agents associated with tube type. Between venipunctures significant variation was found in glucose, alkaline phosphatase and potassium. The authors note that small physiological changes such as arm position, blood flow, and blood pressure as well as changes between venipunctures in depth, tourniquet strength, and duration may all be responsible. While measurement repeatability was the lowest achievable, it was the largest contributor to standard deviation for gamma-glutamyltransferase, alanine aminotransferase and magnesium. The authors conclude that there are many confounding factors to consider when analyzing serum analytes including, those of preprocessing (tube type, clotting time, coagulant), collection (arm position, tourniquet strength, venipuncture depth and duration) and processing (extent of hemolysis, mixing time, time before analysis, and room temperature and humidity) as well as error introduced in measurement.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    LipidClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    ProteinHematology/ auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnatomical location of blood draw
    Arm
    Left and right sides compared
    Biospecimen AcquisitionType of collection container/solution
    Rapid serum tubes (4.0)(RST)
    Serum-separation tubes (SST)-II advance
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageTime at room temperature
    30-33 min
    5-8 min